Perfusion Measurement
Range: 0–200 ml/100g–min
Resolution: < 0.2 ml/100g–min
Accuracy: 10% of full scale (200ml/100g–min)
Volume of Measurement region: ≈ 0.3 ml
Temperature Measurement
Range: 25–46°C
Accuracy: ~ 0.3°C
Resolution: 0.005°C
Hardware Specifications
Power Requirements: US standard 100–120 VAC/60 Hz or 220–240 VAC/50 Hz
Analog Output
Voltage Output: 0 to 2 VDC
Output Scale: Fixed scale 100 ml/100g–min perfusion per volt (0 to 200 ml/100g–min)
Digital Output
Update Period: 1 second
Serial Port: supports data uploading and streaming via RS-232 communication
Electrical Safety
Breakdown Voltage: Medical Grade Isolation: Dielectric strength tested to 4000 VAC
Leakage Current: < 10 µA–Meets IEC-60601 specifications for CF equipment
Physical Specifications
Dimensions: 10.5 L x 10 W x 16.5 H in, 26.67 L x 25.4 W x 41.91 H cm
Weight: 13 lbs (5.9 kg)
Shelf or pole mount
|